Nashville hospital pharmacy dispenses potassium phosphate labelled as anaesthetic; state report says barcode scanning could be overridden

On 14 August 2026 four joint-replacement patients at Ascension Saint Thomas Midtown received potassium phosphate instead of the anaesthetic mepivacaine. FOX 17 News, in a report published on NewsChannel 9's website, said a state survey report found that the pharmacy system let technicians override barcode scanning and that the pharmacist's final check also missed the wrong drug. In November 2026 one patient sued the hospital.

Fact-checked 18 November 2026

TierThe pharmacy systems stayed up. According to the state report as described by FOX 17 News, the barcode check could be overridden with manual entry, which is automation bypassed. How tiers are assigned.

What happened

Ascension Saint Thomas said that on 14 August 2026 four joint-replacement patients at its Midtown campus 'mistakenly received potassium phosphate, instead of the intended anesthetic medication, mepivacaine'. It said leadership self-reported the event to state regulators that day. Citing affiliate WTVF, CBS News reported that the operations were meant to be outpatient procedures. It also reported that the state's healthcare facilities commission alerted the Tennessee Bureau of Investigation, which was investigating. FOX 17 News (WZTV), in a report published on NewsChannel 9's website, described an 80-page Tennessee Health Facilities Commission report, shared with CMS, from a survey completed on 21 August. As FOX 17 describes the report, surveillance video showed a pharmacy technician preparing syringes with potassium phosphate and labelling them as mepivacaine. The report found that the pharmacy system let technicians override barcode scanning and enter medication information manually, and that the pharmacist's final verification did not catch the wrong drug. It also cited a technician with documented training concerns and problems with peer review and staff training, and it placed three areas under 'Immediate Jeopardy'. Ascension said it now stores high-alert medications separately and requires any spinal-medication scan alert to be stopped and resolved through independent verification. A second trained pharmacist now does the final check on spinal medicines. Ascension said CMS removed the Immediate Jeopardy status on 16 September 2026. It also stated: 'We have seen some inaccurate interpretations of this report in the media.' WSMV reported that on 6 November 2026 a 78-year-old patient sued Ascension Saint Thomas in Davidson County Circuit Court, seeking $150 million ($50 million compensatory and $100 million punitive). The complaint alleges that concentrated potassium phosphate was loaded into syringes labelled as spinal anaesthesia, and it claims negligence, recklessness and medical battery. These are allegations and have not been decided. Ascension Saint Thomas said it is cooperating with the lawsuit and will not discuss details while the case is active. The sources reviewed report no outcome of the TBI investigation.[1,2,3,4]

Documented harm

Ascension says the four patients had adverse reactions and received immediate care. FOX 17 News says the state report describes worsening neurological, breathing and blood-pressure problems. CBS News reported families telling WTVF that two patients were left paralysed and one was in intensive care.

What it teaches

  • A barcode check that staff can override with manual entry protects only as well as overrides are restricted and reviewed.
  • When an automated check is bypassed, the next human check carries the whole load, and a visual final check can miss a look-alike vial.
  • Hard stops on scan alerts for high-alert routes such as spinal medication turn an advisory check into a required one.

Sources

  1. Drug mix-up at Nashville hospital causes dire complications in 4 patients. CBS News, 21 August 2026. Secondary Journalism · link checked 2026-11-02
  2. State finds multiple safety failures after Ascension Saint Thomas medication mix-up. FOX 17 News (WZTV), published on newschannel9.com. Supporting Journalism · link checked 2026-11-02
  3. Ascension Saint Thomas Midtown Update. Ascension, September 2026. Supporting Official report · link checked 2026-11-02
  4. 78-year-old victim sues Ascension St. Thomas for damages, here's why her gains may be limited. WSMV, 11 November 2026. Supporting Journalism · link checked 2026-11-28

Pharmacists accept most wrong LLM recommendations when verifying medication orders

In a randomized simulation trial at UCSF, 77 pharmacists and pharmacy trainees verified medication orders with and without GPT-4o recommendations. Assistance raised overall accuracy, but participants accepted 73.9% of incorrect AI recommendations, and their accuracy fell to 26.1% when the AI was wrong.

ICU patient dies after alarms go unanswered for an hour

A 60-year-old ICU patient's monitor alarmed for rising heart rate and falling oxygen saturation; staff responded only after about an hour, when he had stopped breathing.

All incidents · Human handoff

Information only, not advice. FailSystems is an aggregation and synthesis of published sources. It is not consulting, engineering, legal, regulatory or medical advice, and using it creates no professional relationship. Health systems are complex and no approach fits every organisation: anything you adopt is your own decision, at your own risk, and should be checked against the current official sources and by qualified people who know your setting. Full disclaimer.

Dealing with an incident right now? This site is a reference, not an incident-response service. Activate your organisation's emergency operations plan and incident command, and: