Incident

Patient dies while a heart monitor's crisis alarm is switched off

A patient on a cardiac monitor died after the monitor's crisis alarm had been left off; lower-level alarms sounded at the nurses' station but went unheeded.

Fact-checked 18 November 2026

LimitationBoth sources are Boston Globe reports. The state investigators' findings are known here only as the Globe quotes them; no CMS or state inspection record was found online.

TierNo loss of automation: the monitor worked, but its crisis alarm was switched off and lower-level alarms went unheeded. How tiers are assigned.

What happened

The Boston Globe reported in February 2010 that a Massachusetts General Hospital patient had died the previous month after the crisis alarm on a GE heart monitor was inadvertently left off, delaying the response of nurses and doctors. Hospital leaders said they did not know why the alarm was off; a nurse found the patient in crisis on a routine check. Within a day the hospital inspected and disabled the off switch on all 1,100 of its heart monitors, and temporarily assigned a nurse on each unit to listen for alarms. A year later the Globe reported that ten nurses on duty that morning could not recall hearing the lower-level alarms at the central station, or seeing warnings on three hallway signs, as the patient's heart rate fell over about 20 minutes. It quoted state investigators: nursing staff said they were experiencing alarm fatigue and a desensitization to alarms. The investigators did not say why the crisis alarm was off.[1,2]

Documented harm

Death of the patient (Boston Globe).

What it teaches

  • An alarm that can be switched off will sometimes be left off; decide who may silence which alarms, and audit it.
  • Lower-level alarms that sound constantly stop being heard; alarm volume and priority are a safety design problem, not only a staffing one.

Sources

  1. MGH death spurs review of patient monitors. The Boston Globe (Liz Kowalczyk), 21 February 2010. Secondary Journalism · link checked 2026-11-02
  2. Patient alarms often unheard, unheeded. The Boston Globe (Liz Kowalczyk), 13 February 2011. Secondary Journalism · link checked 2026-11-02

ICU patient dies after alarms go unanswered for an hour

A 60-year-old ICU patient's monitor alarmed for rising heart rate and falling oxygen saturation; staff responded only after about an hour, when he had stopped breathing.

Pharmacists accept most wrong LLM recommendations when verifying medication orders

In a randomized simulation trial at UCSF, 77 pharmacists and pharmacy trainees verified medication orders with and without GPT-4o recommendations. Assistance raised overall accuracy, but participants accepted 73.9% of incorrect AI recommendations, and their accuracy fell to 26.1% when the AI was wrong.

Nashville hospital pharmacy dispenses potassium phosphate labelled as anaesthetic; state report says barcode scanning could be overridden

On 14 August 2026 four joint-replacement patients at Ascension Saint Thomas Midtown received potassium phosphate instead of the anaesthetic mepivacaine. FOX 17 News, in a report published on NewsChannel 9's website, said a state survey report found that the pharmacy system let technicians override barcode scanning and that the pharmacist's final check also missed the wrong drug. In November 2026 one patient sued the hospital.

All incidents · Human handoff · Devices

Information only, not advice. FailSystems is an aggregation and synthesis of published sources. It is not consulting, engineering, legal, regulatory or medical advice, and using it creates no professional relationship. Health systems are complex and no approach fits every organisation: anything you adopt is your own decision, at your own risk, and should be checked against the current official sources and by qualified people who know your setting. Full disclaimer.

Dealing with an incident right now? This site is a reference, not an incident-response service. Activate your organisation's emergency operations plan and incident command, and: