how automated healthcare fails, how you'd know, and what to do at each tier — every claim sourced, reviewed continuously


Evidence

Incidents

89 documented failures, each tagged with the layers it crossed and the lowest degradation tier care actually fell to.

2026

Main generator at Gaza's Al-Rantisi children's hospital stops; Health Ministry says vital services halted and a smaller generator cannot cover intensive care

On 25 November 2026 the Ministry of Health in Gaza said the main generator at Al-Rantisi Children's Hospital had stopped suddenly, halting a number of vital medical services and, according to Anadolu's Arabic report, the operating rooms of the neighbouring Eye Hospital. The ministry said repairs were hampered by a complete absence of oils and spare parts.

IT outage takes down the whole IT system at Luxembourg's Centre Hospitalier du Nord; records go to paper and ambulances are redirected

On 23 November 2026 an IT outage affected the entire IT system of the Centre Hospitalier du Nord in Ettelbruck, according to the hospital. Inpatient records were made available on paper, outpatient appointments were cancelled and ambulances were directed to other hospitals until the system was restored shortly after 18:00. The hospital said the cause was purely technical and not a cyberattack.

Suspected ransomware attack on Guadeloupe's university hospital: records unreachable, staff on pen and paper, radiotherapy suspended

On the evening of Saturday 21 November 2026 a cyberattack disrupted IT and telephone systems at the University Hospital of Guadeloupe, which triggered its Plan Blanc emergency plan the next day and postponed non-urgent surgery, consultations and scheduled imaging through 27 November. France's cybersecurity agency ANSSI said the incident appeared to involve ransomware, and the Paris prosecutor opened a criminal investigation.

Cyberattack closes the emergency department at Kreiskrankenhaus Greiz-Ronneburg; emergency calls routed to other hospitals

After a cyberattack was detected on 15 November 2026 at Kreiskrankenhaus Greiz-Ronneburg, the hospital temporarily closed its emergency department. People with emergencies were told to call 112, and the Gera dispatch centre decided where they would be treated. KOMMUNAL reported that the emergency department began taking patients again, step by step, on the evening of 16 November.

Microsoft 365 outage disrupts health-service email and messaging in several countries; NHS falls back to phones and paper where possible

A Microsoft 365 outage that began on 12 November 2026 and lasted more than 24 hours disrupted email and related services for health organisations in several countries, according to CBS News and ABC News (Australia). NHS England activated contingency plans that reverted to other IT systems or paper records, ABC News reported, citing AFP. London hospital trusts said services and procedures were continuing and asked patients to phone instead of emailing.

Network failure at UZ Brussel takes down its shared patient record at partner hospitals; care moves to paper and ambulances are diverted for about a day

A network problem at UZ Brussel affected its IT infrastructure and telephony, according to the hospital. For about 24 hours, planned care was postponed, ambulances were diverted and staff worked on paper. Partner hospitals that use the same UZ Brussel-developed electronic patient record also lost online access to patient files.

PathConnectivity & data → Human handoff

Server outage at Monadnock Community Hospital: ambulance diversion, cancelled elective surgery, lab and radiology, and a week on downtime procedures

An outage that began on 9 November 2026 and that the hospital later called a server outage put Monadnock Community Hospital's emergency department on ambulance diversion. It also led the hospital to cancel elective surgery and outpatient laboratory and radiology appointments. The hospital said it was working under downtime procedures. As of 15 November its phones were restored, other systems were still down, and on 16 November it confirmed that a cyber incident caused the outage (DysruptionHub).

PathConnectivity & data → Human handoff

Study of 2017-2025 FDA data: neurosurgical software recalls, mostly Class I, with adverse-event reports continuing after recall

A peer-reviewed analysis in Neurosurgical Focus identified 20 software-associated neurological device recalls initiated in the US from 2017 to 2025, more than half of them FDA Class I. It counted 16,333 adverse-event notifications for these devices, more than 10,000 of them reported after formal recall notification for two devices.

Cybersecurity event at UNM Health diverts ambulances and cancels clinic appointments and outpatient surgeries for a day

On 26 October 2026 a cybersecurity event at the University of New Mexico Health Sciences Center led UNM Health to divert ambulance traffic to other area health systems for about two hours. It also suspended outpatient surgeries and appointments for the rest of the day. UNM said critical systems were restored in phases from that afternoon, and clinics resumed scheduled care on 27 October.

Lexington, Massachusetts nursing home evacuated after recurring electrical failure; building de-energised

On the night of 24 October 2026 more than 100 residents were evacuated from a nursing and rehabilitation centre in Lexington, Massachusetts, after electrical problems that had recurred over several days. Officials ordered the building de-energised because other building systems, including sewage, were tied into the electrical system. Residents had returned by mid-November, Boston 25 News reported.

University Hospitals in Cleveland disconnects from the internet after unusual server activity; MyChart offline for over a week and some procedures rescheduled

University Hospitals (UH) detected unusual activity on its servers on 22 October 2026 and disconnected its systems from the internet as a precaution. MyChart and other digital services were still offline on 2 November, with no timeline for restoration. UH said it had rescheduled some appointments and elective procedures, but that the vast majority continued as normal.

Quest Diagnostics shuts down its Quanum LSM lab ordering and results system after unauthorized access; clients told to use paper requisitions

After a cyber intrusion, Quest Diagnostics shut down Quanum LSM, the web system some clients use to order lab tests and receive results, and paused two other systems as a precaution. Clients were told to use paper requisitions or another Quest platform. Quest says its ability to serve patients was not materially disrupted.

PathConnectivity & data → Human handoff

Power interruptions at the SPMS data centre take Portugal's national e-prescribing and electronic health record offline

On 19 October 2026, two successive power interruptions at the data centre run by SPMS, the Ministry of Health's shared-services body, left SNS information systems unavailable for several hours. Doctors could not issue electronic prescriptions or consult electronic records, and pharmacies had limited access to prescriptions.

PathPower → Connectivity & data

dcm4chee-arc-light imaging archive: CISA advisory on image deletion, forged studies and unauthenticated patient merges

CISA disclosed six vulnerabilities in the dcm4chee-arc-light medical imaging archive. According to the advisory, they could let an attacker delete stored images, file forged studies, reassign studies to another patient or cause a denial of service. CISA reported no known exploitation and recommends updating to version 5.35.2 or later.

Missed security update to a shared DNS service disrupts all three Singapore public healthcare clusters for about nine and a half hours

On 13 October 2026 a disruption across Singapore's three public healthcare clusters (NHG Health, NUHS and SingHealth) affected appointment booking, visitor registration, medication dispensing and billing for about nine and a half hours. In its reply to parliamentary questions, the Ministry of Health said the cause was a required security update to a shared Domain Name System (DNS) service. Synapxe had contracted a service provider to do the update, and it was not done before a scheduled global change.

PathConnectivity & data → Human handoff

Two power outages knock out Kemptville hospital's EHR, internet and cellular service and their backups; ambulances diverted

On 13 October 2026 two power outages knocked out Kemptville District Hospital's cellular service, internet, electronic health record and its downtime and backup systems. The hospital declared a code grey, moved to paper, asked lower-acuity patients to go elsewhere and diverted ambulances.

PathPower → Connectivity & data → Human handoff

Network outage takes computers offline across Mass General Brigham; hospitals move to downtime procedures

On 8 October 2026 a network outage took some computers and digital devices offline across Mass General Brigham. Its hospitals moved to downtime procedures, including paper forms, and some elective outpatient procedures did not go ahead. MGB said the outage was not linked to a cyberattack; the cause had not been stated.

Luna G3 BPAP devices record the set pressure instead of the pressure delivered when Targeted VT is on; Class I recall

FDA classified a recall of the React Health (3B Medical) Luna G3 BPAP System, model LG3800-30VT, as Class 1. With Targeted VT enabled, the device records and transmits the configured inspiratory pressure instead of the pressure actually delivered. React Health says this could lead clinicians to adjust settings based on inaccurate data, which could delay therapy.

GP website and online-consultation supplier Silicon Practice enters administration; affected practices in England and Wales lose online request channels

Schappit Ltd, trading as Silicon Practice, entered administration in October 2026, and websites, online consultations and chatbots stopped at some of the GP practices it served. NHS England said fewer than 10% of England's roughly 6,000 practices were affected and told patients to contact their practice by phone or in person.

Prague Central Military Hospital: power cut and failed backup pause operations

Doctors told Seznam Zprávy that a power cut at Prague's Central Military Hospital, which backup generators did not adequately cover, paused all operations for about 20 minutes and left ventilators running on internal batteries. The hospital said the interruption lasted several minutes and did not affect patients' health.

Tandem Mobi insulin pumps: false occlusion alarms suspend insulin delivery; Class 1 correction

Tandem Diabetes Care began a correction of Tandem Mobi insulin pumps in September 2026 because the pumps can raise false-positive occlusion alarms that suspend insulin delivery; FDA lists it as a Class 1 recall. A company spokesperson reported 230 adverse events requiring medical intervention or hospitalization as of 29 October 2026, and no deaths.

Manzanillo maternity hospital's backup microsystem and generator fail together during Cuban grid outage

During a national grid outage in Cuba in September 2026, the Fe del Valle Ramos maternity hospital in Manzanillo said its electrical microsystem failed at the same time as its generator. The hospital said the situation was resolved in a short period and that no complications among pregnant women, postpartum women or newborns were recorded.

VOCSN ventilators: loose internal flow-transducer connector can cause reboots; FDA early alert and Class I recall

React Health (Ventec Life Systems) warned that the internal flow transducer cable connector in VOCSN ventilators can unseat and cause reboots and recurring System Fault 10 alarms, which could interrupt ventilation. FDA called it a potentially high-risk device issue, and the VOCSN+Pro entry is classified Class I.

Philips Lumify ultrasound demands online re-registration and blocks scanning in pre-hospital settings

FDA issued an Early Alert on 9 October 2026 and on 19 October 2026 said the issue had been classified as a Class I recall. Philips had reported that its Lumify ultrasound app may stop recognising a registered transducer and block scanning until the transducer is re-registered, which requires Wi-Fi or cellular access. As of 29 September 2026 Philips had reported eight serious injuries and four deaths associated with the issue.

PathConnectivity & data → Devices

Pharmacists accept most wrong LLM recommendations when verifying medication orders

In a randomized simulation trial at UCSF, 77 pharmacists and pharmacy trainees verified medication orders with and without GPT-4o recommendations. Assistance raised overall accuracy, but participants accepted 73.9% of incorrect AI recommendations, and their accuracy fell to 26.1% when the AI was wrong.

Luminis Health cyberattack: ambulance reroutes, paper records and about six weeks of system restoration

A cyberattack on Luminis Health took some systems offline. Ambulances carrying noncritical patients were rerouted away from both of its hospitals' emergency departments, and staff used paper medical records until systems were restored about six weeks later.

Nashville hospital pharmacy dispenses potassium phosphate labelled as anaesthetic; state report says barcode scanning could be overridden

On 14 August 2026 four joint-replacement patients at Ascension Saint Thomas Midtown received potassium phosphate instead of the anaesthetic mepivacaine. FOX 17 News, in a report published on NewsChannel 9's website, said a state survey report found that the pharmacy system let technicians override barcode scanning and that the pharmacist's final check also missed the wrong drug. In November 2026 one patient sued the hospital.

Cedar County Memorial Hospital shuts down its network; EHR offline two weeks and critical imaging patients diverted

After an IT disruption began on 14 August 2026, Cedar County Memorial Hospital shut down its networks and ran all departments on downtime procedures, according to the hospital's incident updates. Its Meditech EHR stayed offline until 28 August, and while imaging could not reach radiologists the emergency department diverted trauma and critical patients who needed immediate advanced imaging.

LATITUDE Clarity server software can hold back AF Burden alerts from LUX-Dx implantable cardiac monitors

Boston Scientific notified clinics that its LATITUDE Clarity remote-monitoring server can delay AF Burden alerts from LUX-Dx insertable cardiac monitors until the next full device interrogation when a non-default threshold is set. The implanted monitors keep detecting and transmitting data; one complaint and no injuries were reported.

Automated Impella Controllers recalled twice for purge cassette recognition failures; Abiomed reported 37 serious injuries and three deaths for the first issue

In 2026 Abiomed twice recalled its Automated Impella Controller, the console that runs Impella heart pumps, over failures to recognise the purge cassette. FDA classified both actions as Class I. FDA says Abiomed reported 37 serious injuries and three deaths linked to the first issue as of 17 August 2026.

Philips PIC iX central monitoring stations can freeze when the watchdog recovery fails; central alarms may be lost

Philips issued a correction for Patient Information Center iX versions 4.0–4.5 after receiving reports that some systems froze and could not be used. According to Health Canada and Philips' notice, a failed watchdog recovery can leave the system in an unknown state in which central monitoring and alarms may not work.

StatStrip Gen 2.0 hospital glucose meters can file a result under the previously scanned patient; Class I recall

Nova Biomedical's StatStrip Gen 2.0 hospital glucose meters can keep a previously scanned patient or quality-control barcode on screen. If the operator accepts it, the glucose result is filed under the wrong patient. FDA classified the recall as Class I, and as of 5 October 2026 Nova had reported one serious injury and no deaths.

Spacelabs Xhibit Central Station: DisplayPort splitter can cause loss of alarm audio and video; Class 2 correction

On 4 June 2026 Spacelabs Healthcare told customers to stop using a DisplayPort splitter (PN 010-1975-00) with its Xhibit Central Station patient-monitoring system. According to the company's field safety notice, the splitter can cause degraded or complete loss of alarm audio and video at the central station or extended displays. FDA lists the action as a Class 2 recall (Z-2916-2026).

Case report: chatbot affirmed a manic patient's delusions and discouraged his antipsychotic; care plan restricted its use

A BMC Psychiatry case report describes a man in his 30s with substance-induced manic psychosis whose ChatGPT conversations, according to the authors, affirmed his perceived 'spiritual awakening' and discouraged his prescribed antipsychotic. His inpatient care plan restricted AI chatbot use; the authors state that any contribution of the chatbot to his illness cannot be determined.

Basement flood cuts power at Jefferson Methodist Hospital, Philadelphia; hospital evacuated

Flooding in the basement of Jefferson Methodist Hospital caused a power outage on the evening of 19 May 2026, and the hospital was evacuated overnight. A police official told NBC10 that the backup generators had failed. The hospital's CEO described running on partial commercial and partial generator power.

Pharmacy AI phone agent pulled from inbound calls after customers reported garbled drug names and wrong refills

Customers told reporters that a pharmacy chain's AI voice and text refill assistant garbled medication names, ordered wrong dosages and unneeded refills, and offered no keypad option. After complaints, which WCAX's headline put in the hundreds, the chain moved inbound calls back to its touch-tone system.

Ransomware at ChipSoft, supplier of patient-record software to most Dutch hospitals, takes hosted portals and links offline

A ransomware attack on ChipSoft, whose software the Dutch government says is used by about 70% of Dutch hospitals, led the company to disconnect the patient portals it hosts and led hospitals to cut connections to it. The government told parliament that care continued, but appointment booking, patients' access to their records and digital referrals moved to phone and email.

Cybersecurity incident at Brockton Hospital: downtime procedures, nine days of ambulance diversion, chemotherapy and pharmacy disruption

Signature Healthcare Brockton Hospital moved to downtime procedures and diverted ambulances after finding suspicious activity on its network on 6 April 2026. Chemotherapy infusions were cancelled and its retail pharmacies could not fill prescriptions. The hospital's alerts page says diversion ended on 15 April and full pharmacy operations resumed on 24 April.

Ransomware at AZ Monica, Antwerp: servers shut down, critical patients transferred, three weeks on paper

AZ Monica shut down all servers at its Antwerp and Deurne campuses after a cyberattack detected early on 13 January 2026. Scheduled surgery stopped, seven critical-care patients were transferred, and according to the hospital's CEO staff worked mainly on paper until the electronic patient record was restarted three weeks later.

Medicare's AI-assisted WISeR prior-authorization pilot: records and reporting describe long delays and erroneous denials

CMS's WISeR model, in which technology vendors use AI and machine learning to review prior-authorization requests in traditional Medicare, launched in January 2026 in six states. CMS records obtained by EFF and reporting by KFF Health News describe long delays, an erroneous denial and a vendor that said it would auto-approve requests until its system was fully functional.

2025

FreeStyle Libre 3 sensors report falsely low glucose

Some Libre 3 and 3 Plus continuous glucose sensors read lower than actual glucose; FDA classified the correction Class I after reports of 860 serious injuries and 7 deaths.

Nihon Kohden CNS-6201 central monitors: unpatched denial-of-service flaw in an end-of-support product

JVN and CISA disclosed that a crafted network packet can make Nihon Kohden's CNS-6201 central monitor terminate abnormally; no patch is offered because the product is out of support. In September 2026 the firm began a Class 2 correction, saying exploitation could disrupt centralised monitoring of multiple beds and the display of alarms.

Baxter Novum IQ infusion pumps: underinfusion and software faults lead to Class I corrections and a distribution hold

FDA issued early alerts in July and August 2025 about Baxter's Novum IQ large volume pump, which can underinfuse after certain rate increases, and about software faults in both Novum IQ pumps. The corrections were classified as Class I recalls. Baxter states that it temporarily stopped distributing and installing the large volume pump in the US and Canada in July 2025, except in cases of medical necessity.

Faulty extension cord cuts power to Sabah hospital operating theatres; UPS fault keeps the generator from taking over

In a statement on 19 November 2026, the Sabah Health Department (JKNS) said a faulty extension cord caused a power disruption in the operating theatres of Sabah Women and Children's Hospital on 13 May 2025. It said the UPS did not work and the emergency generator did not take over.

2024

UMC Health System in Lubbock diverts ambulances and runs clinics on downtime procedures after ransomware attack

A ransomware attack on 26 September 2024 caused an IT outage at UMC Health System in Lubbock. Ambulances carrying emergency and non-emergency patients were diverted to nearby facilities. Clinics stayed open on downtime procedures while doctors lacked access to many patient records.

CrowdStrike Falcon content update crashes Windows hosts, including hospital systems

A faulty Rapid Response Content update to CrowdStrike's Falcon sensor crashed about 8.5 million Windows devices worldwide. Outside-in measurement found disrupted services at 759 of 2,232 US hospitals studied.

PathDevices → Connectivity & data → Human handoff

Synnovis pathology ransomware, South-East London

Ransomware hit Synnovis, the pathology provider for several south-east London NHS trusts and GP practices. Blood testing and matching collapsed, more than 11,000 appointments and procedures were postponed, O-type blood ran short nationally, and one death was later partly attributed to a delayed result.

PathConnectivity & data → Human handoff

Ascension ransomware and multi-week EHR downtime

A ransomware attack took Ascension's electronic records offline for about five weeks. Clinicians told KFF Health News of medication errors and delayed lab results, and one said he had no training for the attack; Ascension said its care teams were trained for such disruptions.

Change Healthcare ransomware and national claims/pharmacy clearinghouse outage

Attackers used stolen credentials on a Change Healthcare Citrix remote-access portal that had no multi-factor authentication, then deployed ransomware nine days later. Disconnecting the clearinghouse stalled pharmacy claims, medical claims and payments across the US.

PathConnectivity & data → Human handoff

2023

2022

2021

Ransomware spillover to adjacent San Diego emergency departments

A month-long ransomware attack on a health system with about 25% of regional inpatient discharges was associated with higher census, ambulance arrivals, waits and stroke activations at two unaffected academic EDs nearby, according to a before-and-after cohort study.

PathConnectivity & data → Human handoff

2020

2019

2018

2017

2013

2012

2010

ICU patient dies after alarms go unanswered for an hour

A 60-year-old ICU patient's monitor alarmed for rising heart rate and falling oxygen saturation; staff responded only after about an hour, when he had stopped breathing.

2009

2005

2003

2002

Information only, not advice. FailSystems is an aggregation and synthesis of published sources. It is not consulting, engineering, legal, regulatory or medical advice, and using it creates no professional relationship. Health systems are complex and no approach fits every organisation: anything you adopt is your own decision, at your own risk, and should be checked against the current official sources and by qualified people who know your setting. Full disclaimer.