Incident

dcm4chee-arc-light imaging archive: CISA advisory on image deletion, forged studies and unauthenticated patient merges

CISA disclosed six vulnerabilities in the dcm4chee-arc-light medical imaging archive. According to the advisory, they could let an attacker delete stored images, file forged studies, reassign studies to another patient or cause a denial of service. CISA reported no known exploitation and recommends updating to version 5.35.2 or later.

Sources checked when written 30 October 2026

TierThis is a vulnerability disclosure with no reported exploitation, and no disruption of care is documented. How tiers are assigned.

What happened

On 15 October 2026 CISA published medical advisory ICSMA-26-288-02, which describes six vulnerabilities in J4Care dcm4chee-arc-light 5.35.1. The highest CVSS v3.1 score is 9.1. According to CISA, a REST service for changing storage status lets location records linked to DICOM instances be set to deletion states, which deletes the referenced objects from storage. The default configuration accepts DICOM associations from unregistered calling AEs, which allows retrieval of patient data and images and filing of fabricated studies. The default HL7 listener accepts ADT^A40 patient-merge messages without authenticating the sender, so studies can be reassigned to another patient, and an unmerge does not reverse this. Three parsing flaws, in the JPEG, MP4 and multipart input handlers, can cause denial of service. CISA said no known public exploitation had been reported to it. It credited researcher Abhinav Agarwal and recommended upgrading to 5.35.2 or later and keeping imaging systems off the internet, behind firewalls. J4Care's release notes for version 5.35.2 on GitHub, dated 15 October, list security fixes for mass deletion of stored objects through an unrestricted storage status change and for the three parsing denial-of-service issues. The release notes do not mention the calling-AE or HL7 merge findings.[1,2]

Documented harm

none documented

What it teaches

  • Imaging archives that accept DICOM or HL7 traffic from any sender by default should be limited to known peers when they are installed, not after an advisory.
  • Patient-merge messages change which patient a study belongs to, and where a merge cannot be cleanly reversed it needs sender authentication and an audit trail.
  • Integrity attacks such as deleted or forged images and reassigned studies may not cause any outage, so detecting them depends on reconciliation and audit rather than availability monitoring.

Sources

  1. ICS Medical Advisory ICSMA-26-288-02: J4Care dcm4chee-arc-light. Cybersecurity and Infrastructure Security Agency, 15 October 2026. Primary Official report · link checked 2026-11-02
  2. dcm4chee-arc-light 5.35.2 release notes. J4Care / dcm4che project (GitHub), 15 October 2026. Supporting Guidance · link checked 2026-11-02

Philips Lumify ultrasound demands online re-registration and blocks scanning in pre-hospital settings

FDA issued an Early Alert on 9 October 2026 and on 19 October 2026 said the issue had been classified as a Class I recall. Philips had reported that its Lumify ultrasound app may stop recognising a registered transducer and block scanning until the transducer is re-registered, which requires Wi-Fi or cellular access. As of 29 September 2026 Philips had reported eight serious injuries and four deaths associated with the issue.

PathConnectivity & data → Devices

Nihon Kohden CNS-6201 central monitors: unpatched denial-of-service flaw in an end-of-support product

JVN and CISA disclosed that a crafted network packet can make Nihon Kohden's CNS-6201 central monitor terminate abnormally; no patch is offered because the product is out of support. In September 2026 the firm began a Class 2 correction, saying exploitation could disrupt centralised monitoring of multiple beds and the display of alarms.

CrowdStrike Falcon content update crashes Windows hosts, including hospital systems

A faulty Rapid Response Content update to CrowdStrike's Falcon sensor crashed about 8.5 million Windows devices worldwide. Outside-in measurement found disrupted services at 759 of 2,232 US hospitals studied.

PathDevices → Connectivity & data → Human handoff

Study of 2017-2025 FDA data: neurosurgical software recalls, mostly Class I, with adverse-event reports continuing after recall

A peer-reviewed analysis in Neurosurgical Focus identified 20 software-associated neurological device recalls initiated in the US from 2017 to 2025, more than half of them FDA Class I. It counted 16,333 adverse-event notifications for these devices, more than 10,000 of them reported after formal recall notification for two devices.

All incidents · Devices · Connectivity & data

Information only, not advice. FailSystems is an aggregation and synthesis of published sources. It is not consulting, engineering, legal, regulatory or medical advice, and using it creates no professional relationship. Health systems are complex and no approach fits every organisation: anything you adopt is your own decision, at your own risk, and should be checked against the current official sources and by qualified people who know your setting. Full disclaimer.