On 15 October 2026 CISA published medical advisory ICSMA-26-288-02, which describes six vulnerabilities in J4Care dcm4chee-arc-light 5.35.1. The highest CVSS v3.1 score is 9.1. According to CISA, a REST service for changing storage status lets location records linked to DICOM instances be set to deletion states, which deletes the referenced objects from storage. The default configuration accepts DICOM associations from unregistered calling AEs, which allows retrieval of patient data and images and filing of fabricated studies. The default HL7 listener accepts ADT^A40 patient-merge messages without authenticating the sender, so studies can be reassigned to another patient, and an unmerge does not reverse this. Three parsing flaws, in the JPEG, MP4 and multipart input handlers, can cause denial of service. CISA said no known public exploitation had been reported to it. It credited researcher Abhinav Agarwal and recommended upgrading to 5.35.2 or later and keeping imaging systems off the internet, behind firewalls.
J4Care's release notes for version 5.35.2 on GitHub, dated 15 October, list security fixes for mass deletion of stored objects through an unrestricted storage status change and for the three parsing denial-of-service issues. The release notes do not mention the calling-AE or HL7 merge findings.[1,2]