According to FDA, a software issue can cause the Philips Lumify Diagnostic Ultrasound System app to stop recognising a previously registered transducer and prompt users to re-register without warning. Scanning is blocked until registration is complete, and registration requires Wi-Fi or cellular signal access. Philips received reports that these re-registration events prevented scanning in pre-hospital environments and delayed care. According to FDA, those events happened outside Wi-Fi or cellular access, and devices used outside the hospital, such as in ambulances, aircraft or field settings, are at greater risk. A separate issue is that the USB connection between the transducer and the phone or tablet may disconnect or become unstable during use.
As of 29 September 2026, Philips had reported eight serious injuries and four deaths associated with the issue, according to FDA. Affected app versions include iOS 2.X, 5.0, 5.1.2 and 5.1.3 and Android 1.X, 3.X, 4.X, 5.1 and 5.1.1. Philips wrote to all affected customers on 24 September 2026. It advised using the system only where Wi-Fi or cellular signal is readily available unless a backup ultrasound device is immediately available. FDA issued an Early Alert on 9 October 2026. On 19 October 2026 FDA updated the communication to say the issue had been classified as a Class I recall, the most serious type, and that the affected products and recommendations had not changed. FDA said Philips is developing a software update and will send a follow-up communication once it is available. A Philips spokesperson told MedTech Dive that the company is also working on a solution to the USB issue and has set up a dedicated taskforce to support affected customers.[1,2,3]