Qureshi and colleagues analysed software-related neurological medical device recalls initiated in the US from 2017 to 2025, drawing on FDA's recall, MAUDE, AccessGUDID and openFDA databases. According to the paper, they identified 20 unique software-associated recalls across 7 companies: surgical navigation 55%, neurophysiological monitoring 25%, surgical planning 15% and stereotactic devices 5%. More than half were Class I, and the annual frequency of Class I recalls rose from 2017 to 2024 (p = 0.027). The authors counted 16,333 adverse-event notifications: 93.4% related to malfunction and 5.9% to injury. Two surgical navigation devices accounted for 88.8% of these notifications, and more than 10,000 were reported after formal recall notification for those two devices. System instability (55.0%) and software defects (43.5%) made up 98.5% of notifications. Severe harm was reported more often for Class II recalls than for Class I recalls (14.6% vs 3.6%, p < 0.001); the authors describe this as a discordance between recall class and reported harm. Much of the adverse-event data comes from MAUDE, whose reports FDA states are not verified and cannot by themselves show that a device caused an event.
FDA's recall notices show the kind of defect in this category. In April 2024 Medtronic began a Class I correction of StealthStation S8 Application versions 1.0.1 to 1.2.0 (876 devices in the US, distributed March 2017 to March 2024). According to FDA, a software error could drop digits or letters from the 'distance to target' or 'tip stop point' display during navigation, so the biopsy depth stop could be set too short or too long. FDA said Medtronic had received 28 complaints and no reports of injury or death, and that the remedy was a free software update.[1,2,3]