Study of 2017-2025 FDA data: neurosurgical software recalls, mostly Class I, with adverse-event reports continuing after recall

A peer-reviewed analysis in Neurosurgical Focus identified 20 software-associated neurological device recalls initiated in the US from 2017 to 2025, more than half of them FDA Class I. It counted 16,333 adverse-event notifications for these devices, more than 10,000 of them reported after formal recall notification for two devices.

Sources checked when written 28 November 2026

TierStudy; tiers apply to events only. How tiers are assigned.

What happened

Qureshi and colleagues analysed software-related neurological medical device recalls initiated in the US from 2017 to 2025, drawing on FDA's recall, MAUDE, AccessGUDID and openFDA databases. According to the paper, they identified 20 unique software-associated recalls across 7 companies: surgical navigation 55%, neurophysiological monitoring 25%, surgical planning 15% and stereotactic devices 5%. More than half were Class I, and the annual frequency of Class I recalls rose from 2017 to 2024 (p = 0.027). The authors counted 16,333 adverse-event notifications: 93.4% related to malfunction and 5.9% to injury. Two surgical navigation devices accounted for 88.8% of these notifications, and more than 10,000 were reported after formal recall notification for those two devices. System instability (55.0%) and software defects (43.5%) made up 98.5% of notifications. Severe harm was reported more often for Class II recalls than for Class I recalls (14.6% vs 3.6%, p < 0.001); the authors describe this as a discordance between recall class and reported harm. Much of the adverse-event data comes from MAUDE, whose reports FDA states are not verified and cannot by themselves show that a device caused an event. FDA's recall notices show the kind of defect in this category. In April 2024 Medtronic began a Class I correction of StealthStation S8 Application versions 1.0.1 to 1.2.0 (876 devices in the US, distributed March 2017 to March 2024). According to FDA, a software error could drop digits or letters from the 'distance to target' or 'tip stop point' display during navigation, so the biopsy depth stop could be set too short or too long. FDA said Medtronic had received 28 complaints and no reports of injury or death, and that the remedy was a free software update.[1,2,3]

Documented harm

The study reports that 5.9% of 16,333 adverse-event notifications related to injury, and that severe harm was reported in 14.6% of notifications for Class II recalls and 3.6% for Class I. These are unverified reports, not confirmed device-caused harm. For the StealthStation S8 example, FDA reports no injuries or deaths.

What it teaches

  • A recall notice does not end exposure: if reports keep arriving after a recall, the affected software is often still in clinical use, so tracking whether fixes are actually installed matters.
  • Recall class reflects the regulator's assessment of potential risk and may not match the harm reported in the field, so local risk reviews should look at both.
  • Runtime instability, not only wrong calculations, is a major failure mode for surgical software, so procedures relying on navigation need a planned fallback when the system freezes or restarts mid-case.

Sources

  1. Neurosurgical software device safety in the US: analysis of 2017-2025 FDA Recall and Adverse Event Data. Neurosurgical Focus (Journal of Neurosurgery Publishing Group), 1 November 2026. Primary Peer-reviewed · link checked 2026-11-02
  2. Medtronic Recalls StealthStation S8 Application Versions 1.2.0, 1.1.0, 1.0.3, 1.0.2, and 1.0.1 for Software Error that May Cause Incorrect Measurements During Cranial Surgery. U.S. Food and Drug Administration, June 2024. Primary Official report · link checked 2026-11-02
  3. About Manufacturer and User Facility Device Experience (MAUDE) Database. U.S. Food and Drug Administration, 8 December 2025. Primary Dataset · link checked 2026-11-02

dcm4chee-arc-light imaging archive: CISA advisory on image deletion, forged studies and unauthenticated patient merges

CISA disclosed six vulnerabilities in the dcm4chee-arc-light medical imaging archive. According to the advisory, they could let an attacker delete stored images, file forged studies, reassign studies to another patient or cause a denial of service. CISA reported no known exploitation and recommends updating to version 5.35.2 or later.

Luna G3 BPAP devices record the set pressure instead of the pressure delivered when Targeted VT is on; Class I recall

FDA classified a recall of the React Health (3B Medical) Luna G3 BPAP System, model LG3800-30VT, as Class 1. With Targeted VT enabled, the device records and transmits the configured inspiratory pressure instead of the pressure actually delivered. React Health says this could lead clinicians to adjust settings based on inaccurate data, which could delay therapy.

Tandem Mobi insulin pumps: false occlusion alarms suspend insulin delivery; Class 1 correction

Tandem Diabetes Care began a correction of Tandem Mobi insulin pumps in September 2026 because the pumps can raise false-positive occlusion alarms that suspend insulin delivery; FDA lists it as a Class 1 recall. A company spokesperson reported 230 adverse events requiring medical intervention or hospitalization as of 29 October 2026, and no deaths.

VOCSN ventilators: loose internal flow-transducer connector can cause reboots; FDA early alert and Class I recall

React Health (Ventec Life Systems) warned that the internal flow transducer cable connector in VOCSN ventilators can unseat and cause reboots and recurring System Fault 10 alarms, which could interrupt ventilation. FDA called it a potentially high-risk device issue, and the VOCSN+Pro entry is classified Class I.

Philips Lumify ultrasound demands online re-registration and blocks scanning in pre-hospital settings

FDA issued an Early Alert on 9 October 2026 and on 19 October 2026 said the issue had been classified as a Class I recall. Philips had reported that its Lumify ultrasound app may stop recognising a registered transducer and block scanning until the transducer is re-registered, which requires Wi-Fi or cellular access. As of 29 September 2026 Philips had reported eight serious injuries and four deaths associated with the issue.

PathConnectivity & data → Devices

All incidents · Devices

Information only, not advice. FailSystems is an aggregation and synthesis of published sources. It is not consulting, engineering, legal, regulatory or medical advice, and using it creates no professional relationship. Health systems are complex and no approach fits every organisation: anything you adopt is your own decision, at your own risk, and should be checked against the current official sources and by qualified people who know your setting. Full disclaimer.

Dealing with an incident right now? This site is a reference, not an incident-response service. Activate your organisation's emergency operations plan and incident command, and: