Alaris infusion interoperability backlog can load outdated pump orders

A Class I software correction found that backlogged EHR-to-pump automated programming requests could load stale rate, dose or volume parameters.

Sources checked when written 1 November 2026

TierClinicians were told to verify every pump parameter against the order before starting: automation demoted to advice. How tiers are assigned.

What happened

FDA announced that with BD Alaris Systems Manager (v12.5.2 and earlier) and Care Coordination Engine Infusion Adapter (v1.7.2), delayed responses could cause automated programming requests to back up, so an outdated request could later load on the pump with different rate, dose or volume parameters; if accepted, the patient could receive the wrong dose. It was classified as the most serious recall type. No injuries or deaths were reported. Clinicians were told to verify all parameters before starting and to update the adapter software. The same pump platform had a Class I software recall in 2020 for five issues including a low-battery alarm failure.[1,2]

Documented harm

None reported (FDA).

What it teaches

  • Integration queues turn latency into stale commands; a late order is a wrong order.
  • Auto-programming does not remove the need for a human check at the pump; it moves the check.
  • Watch interface engine queue depth as a patient safety signal, not an IT metric.

Sources

  1. Infusion Pump Software Correction: BD Issues Correction for BD Alaris Systems Manager and Care Coordination Engine Infusion Adapter Software Due to Risk for Outdated Automated Programming Requests to Load. U.S. Food and Drug Administration, 18 February 2025. Primary Official report · link checked 2026-11-02
  2. BD Provides Update on Feb. 4, 2020 Voluntary Recall of the BD Alaris System PC Units and Modules. U.S. Food and Drug Administration (company announcement), 9 March 2020. Primary Official report · link checked 2026-11-02

CrowdStrike Falcon content update crashes Windows hosts, including hospital systems

A faulty Rapid Response Content update to CrowdStrike's Falcon sensor crashed about 8.5 million Windows devices worldwide. Outside-in measurement found disrupted services at 759 of 2,232 US hospitals studied.

PathDevices → Connectivity & data → Human handoff

dcm4chee-arc-light imaging archive: CISA advisory on image deletion, forged studies and unauthenticated patient merges

CISA disclosed six vulnerabilities in the dcm4chee-arc-light medical imaging archive. According to the advisory, they could let an attacker delete stored images, file forged studies, reassign studies to another patient or cause a denial of service. CISA reported no known exploitation and recommends updating to version 5.35.2 or later.

Luna G3 BPAP devices record the set pressure instead of the pressure delivered when Targeted VT is on; Class I recall

FDA classified a recall of the React Health (3B Medical) Luna G3 BPAP System, model LG3800-30VT, as Class 1. With Targeted VT enabled, the device records and transmits the configured inspiratory pressure instead of the pressure actually delivered. React Health says this could lead clinicians to adjust settings based on inaccurate data, which could delay therapy.

Philips Lumify ultrasound demands online re-registration and blocks scanning in pre-hospital settings

FDA issued an Early Alert on 9 October 2026 and on 19 October 2026 said the issue had been classified as a Class I recall. Philips had reported that its Lumify ultrasound app may stop recognising a registered transducer and block scanning until the transducer is re-registered, which requires Wi-Fi or cellular access. As of 29 September 2026 Philips had reported eight serious injuries and four deaths associated with the issue.

PathConnectivity & data → Devices

LATITUDE Clarity server software can hold back AF Burden alerts from LUX-Dx implantable cardiac monitors

Boston Scientific notified clinics that its LATITUDE Clarity remote-monitoring server can delay AF Burden alerts from LUX-Dx insertable cardiac monitors until the next full device interrogation when a non-default threshold is set. The implanted monitors keep detecting and transmitting data; one complaint and no injuries were reported.

Philips PIC iX central monitoring stations can freeze when the watchdog recovery fails; central alarms may be lost

Philips issued a correction for Patient Information Center iX versions 4.0–4.5 after receiving reports that some systems froze and could not be used. According to Health Canada and Philips' notice, a failed watchdog recovery can leave the system in an unknown state in which central monitoring and alarms may not work.

All incidents · Devices · Connectivity & data · Human handoff

Information only, not advice. FailSystems is an aggregation and synthesis of published sources. It is not consulting, engineering, legal, regulatory or medical advice, and using it creates no professional relationship. Health systems are complex and no approach fits every organisation: anything you adopt is your own decision, at your own risk, and should be checked against the current official sources and by qualified people who know your setting. Full disclaimer.