In an August 2026 Urgent Medical Device Correction, Boston Scientific said that when the AF Burden alert threshold is set to '>0 hours per day' (a non-default configuration), the LATITUDE Clarity server does not convert the daily device check into a full device interrogation as intended. Without a full interrogation the system cannot evaluate the alert, so notification of an AF Burden alert may be delayed until the next full interrogation, which is triggered by another alert, a scheduled remote follow-up, or a patient- or clinic-initiated interrogation. The company said the implanted monitor keeps detecting AF, storing data and transmitting device data.
Boston Scientific reported one complaint and no patient injuries or deaths, and said a server software update is planned. Until then it advised changing the threshold to another setting or enabling AF Event alerts with a minimum duration such as 2 minutes, and does not recommend replacing implanted monitors. It said delayed notification may delay clinical evaluation or treatment of an atrial arrhythmia, while the most serious consequences remain hypothetical. FDA's recall database lists the correction as Class 2 (Z-3081-2026), initiated 6 August 2026 and posted 2 September 2026, with software design as the cause.[1,2]