On 12 November 2026 FDA posted a Class 1 recall (Z-0409-2027) of the Luna G3 BPAP System, model LG3800-30VT. The FDA record describes it as a bi-level positive airway pressure device for treating adult obstructive sleep apnea. The recalling firm is 3B Medical, Inc. (React Health). The FDA record gives the reason as "The configured IPAP value is recorded and transmitted instead of the actual delivered inspiratory pressure when Targeted VT is enabled", and FDA determined the cause as device design. React Health's notice says that with Targeted VT enabled, affected devices do not record the pressure actually delivered on the internal SD card. They also do not transmit it to the company's cloud-based management system, React Health Connect. Devices without Targeted VT enabled are not affected. According to React Health, the inaccurate pressure data in historical therapy reports "could lead a clinician to adjust device settings based on inaccurate therapy pressure data, which could result in delayed therapy until the setting is corrected."
The FDA record dates the firm's initiation to 8 October 2026 and says the firm began notifying customers on 9 October. React Health dates its original notification 8 October and published an update on 17 November. The two sources give different counts. The FDA record lists 12,672 units in commerce in the US, including Puerto Rico. React Health's notice says 7,533 devices were distributed in the US and Puerto Rico. The FDA record tells customers to:
- discontinue Targeted VT immediately,
- remove affected devices from clinical service,
- use unaffected PAP devices for patient support,
- install a firmware correction from React Health.
React Health's notice tells patients to contact their provider or equipment supplier. It also tells them not to change settings or stop therapy on their own, and to keep using the device as prescribed until they receive guidance. It names firmware G3-2.00.80 as the correction. It says data collected with Targeted VT enabled before the upgrade will not be changed and will still show the previously recorded pressures.[1,2]