FreeStyle Libre 3 sensors report falsely low glucose

Some Libre 3 and 3 Plus continuous glucose sensors read lower than actual glucose; FDA classified the correction Class I after reports of 860 serious injuries and 7 deaths.

Fact-checked 27 November 2026

TierThe automation stayed up and kept producing output that was wrong or went unchecked; there was no outage to fall back from. How tiers are assigned.

What happened

Abbott initiated a correction on 24 November 2025 after finding certain FreeStyle Libre 3 and 3 Plus sensors could give glucose readings lower than actual levels, which could lead users to take excess carbohydrate or skip or delay insulin. FDA classified it Class I. As of 7 January 2026, 860 serious injuries and 7 deaths had been reported. Users were told to check serial numbers, stop using affected sensors and use a blood glucose meter for treatment decisions until replacements arrived.[1,2]

Documented harm

860 serious injuries and 7 deaths reported as of 7 Jan 2026 (FDA).

What it teaches

  • Silent measurement error in a home device reaches patients before it reaches the hospital.
  • The fallback (fingerstick meter) must already be in the patient's hands and they must know when to use it.
  • ECRI ranks recall communication failures for home diabetes technology a top-4 hazard for 2026.

Sources

  1. Glucose Monitor Sensor Recall: Abbott Diabetes Care Removes Certain FreeStyle Libre 3 and FreeStyle Libre 3 Plus Sensors. U.S. Food and Drug Administration, 5 February 2026. Primary Official report · link checked 2026-11-02
  2. Top 10 Health Technology Hazards for 2026: Executive Brief. ECRI, January 2026. Secondary Official report · link checked 2026-11-02

Study of 2017-2025 FDA data: neurosurgical software recalls, mostly Class I, with adverse-event reports continuing after recall

A peer-reviewed analysis in Neurosurgical Focus identified 20 software-associated neurological device recalls initiated in the US from 2017 to 2025, more than half of them FDA Class I. It counted 16,333 adverse-event notifications for these devices, more than 10,000 of them reported after formal recall notification for two devices.

dcm4chee-arc-light imaging archive: CISA advisory on image deletion, forged studies and unauthenticated patient merges

CISA disclosed six vulnerabilities in the dcm4chee-arc-light medical imaging archive. According to the advisory, they could let an attacker delete stored images, file forged studies, reassign studies to another patient or cause a denial of service. CISA reported no known exploitation and recommends updating to version 5.35.2 or later.

Luna G3 BPAP devices record the set pressure instead of the pressure delivered when Targeted VT is on; Class I recall

FDA classified a recall of the React Health (3B Medical) Luna G3 BPAP System, model LG3800-30VT, as Class 1. With Targeted VT enabled, the device records and transmits the configured inspiratory pressure instead of the pressure actually delivered. React Health says this could lead clinicians to adjust settings based on inaccurate data, which could delay therapy.

Tandem Mobi insulin pumps: false occlusion alarms suspend insulin delivery; Class 1 correction

Tandem Diabetes Care began a correction of Tandem Mobi insulin pumps in September 2026 because the pumps can raise false-positive occlusion alarms that suspend insulin delivery; FDA lists it as a Class 1 recall. A company spokesperson reported 230 adverse events requiring medical intervention or hospitalization as of 29 October 2026, and no deaths.

VOCSN ventilators: loose internal flow-transducer connector can cause reboots; FDA early alert and Class I recall

React Health (Ventec Life Systems) warned that the internal flow transducer cable connector in VOCSN ventilators can unseat and cause reboots and recurring System Fault 10 alarms, which could interrupt ventilation. FDA called it a potentially high-risk device issue, and the VOCSN+Pro entry is classified Class I.

All incidents · Devices

Information only, not advice. FailSystems is an aggregation and synthesis of published sources. It is not consulting, engineering, legal, regulatory or medical advice, and using it creates no professional relationship. Health systems are complex and no approach fits every organisation: anything you adopt is your own decision, at your own risk, and should be checked against the current official sources and by qualified people who know your setting. Full disclaimer.