VOCSN ventilators: loose internal flow-transducer connector can cause reboots; FDA early alert and Class I recall

React Health (Ventec Life Systems) warned that the internal flow transducer cable connector in VOCSN ventilators can unseat and cause reboots and recurring System Fault 10 alarms, which could interrupt ventilation. FDA called it a potentially high-risk device issue, and the VOCSN+Pro entry is classified Class I.

Fact-checked 25 November 2026

TierFDA reports no reboots during active ventilation and no serious injuries or deaths, so no drop in care is documented. How tiers are assigned.

What happened

According to FDA's early alert (page current as of 23 September 2026), React Health stated that the internal flow transducer (IFT) cable connector in VOCSN ventilators with serial numbers 112786 and higher may partially or fully unseat, causing loss of communication with the IFT board. This can lead to reboots and recurring System Fault 10 alarms, and dropping the device can cause it. FDA said that a reboot interrupting therapy could delay therapy or cause a serious adverse event. The alert covers VOCSN+Pro, VC+Pro, VOCSN, VCSN+Pro, V+Pro, V+Pro Emergency, VC and V*Home, including Japan variants. FDA said no reboots during active ventilation had been reported, that it considers such a reboot during active therapy 'remote', and that as of 12 August 2026 React Health had reported no serious injuries or deaths. React Health sent its customer letter on 17 September 2026. FDA's recall database lists the VOCSN+Pro (model PRT-00490-001) entry as a Class 1 recall, Z-0122-2027, posted 20 October 2026 and still open. It records 4,369 units in commerce, distributed across the US including Puerto Rico and outside the US to Japan and China; FDA lists the cause as under investigation by the firm. Recommended actions include using backup ventilation as needed, not stopping therapy unless an alternative is available and a clinician instructs it, keeping alternative ventilation equipment ready, continuously monitoring ventilator-dependent patients, with pulse oximetry and CO2 monitoring if available, and performing the checkout procedure after a drop or impact.[1,2,3]

Documented harm

None documented (no serious injuries or deaths reported as of 12 August 2026, per FDA).

What it teaches

  • Portable life-support devices face drops and transport stresses that can loosen internal connections, so post-impact checks belong in routine use.
  • For ventilator-dependent patients at home, a ready backup ventilator or manual resuscitator and a trained caregiver are the real fallback when a device reboots.
  • A recurring fault alarm can be the only early warning of a hardware fault, so caregivers need to know which alarms call for servicing.

Sources

  1. Early Alert: Ventilator Issue from React Health. U.S. Food and Drug Administration, 23 September 2026. Primary Official report · link checked 2026-11-02
  2. Class 1 Device Recall VOCSN+Pro (Z-0122-2027). U.S. Food and Drug Administration (Medical Device Recalls database), 20 October 2026. Primary Dataset · link checked 2026-11-02
  3. Early Alert for React Health VOCSN Devices. Respiratory Therapy, 25 September 2026. Supporting Journalism · link checked 2026-11-02

Study of 2017-2025 FDA data: neurosurgical software recalls, mostly Class I, with adverse-event reports continuing after recall

A peer-reviewed analysis in Neurosurgical Focus identified 20 software-associated neurological device recalls initiated in the US from 2017 to 2025, more than half of them FDA Class I. It counted 16,333 adverse-event notifications for these devices, more than 10,000 of them reported after formal recall notification for two devices.

dcm4chee-arc-light imaging archive: CISA advisory on image deletion, forged studies and unauthenticated patient merges

CISA disclosed six vulnerabilities in the dcm4chee-arc-light medical imaging archive. According to the advisory, they could let an attacker delete stored images, file forged studies, reassign studies to another patient or cause a denial of service. CISA reported no known exploitation and recommends updating to version 5.35.2 or later.

Luna G3 BPAP devices record the set pressure instead of the pressure delivered when Targeted VT is on; Class I recall

FDA classified a recall of the React Health (3B Medical) Luna G3 BPAP System, model LG3800-30VT, as Class 1. With Targeted VT enabled, the device records and transmits the configured inspiratory pressure instead of the pressure actually delivered. React Health says this could lead clinicians to adjust settings based on inaccurate data, which could delay therapy.

Tandem Mobi insulin pumps: false occlusion alarms suspend insulin delivery; Class 1 correction

Tandem Diabetes Care began a correction of Tandem Mobi insulin pumps in September 2026 because the pumps can raise false-positive occlusion alarms that suspend insulin delivery; FDA lists it as a Class 1 recall. A company spokesperson reported 230 adverse events requiring medical intervention or hospitalization as of 29 October 2026, and no deaths.

Philips Lumify ultrasound demands online re-registration and blocks scanning in pre-hospital settings

FDA issued an Early Alert on 9 October 2026 and on 19 October 2026 said the issue had been classified as a Class I recall. Philips had reported that its Lumify ultrasound app may stop recognising a registered transducer and block scanning until the transducer is re-registered, which requires Wi-Fi or cellular access. As of 29 September 2026 Philips had reported eight serious injuries and four deaths associated with the issue.

PathConnectivity & data → Devices

All incidents · Devices

Information only, not advice. FailSystems is an aggregation and synthesis of published sources. It is not consulting, engineering, legal, regulatory or medical advice, and using it creates no professional relationship. Health systems are complex and no approach fits every organisation: anything you adopt is your own decision, at your own risk, and should be checked against the current official sources and by qualified people who know your setting. Full disclaimer.

Dealing with an incident right now? This site is a reference, not an incident-response service. Activate your organisation's emergency operations plan and incident command, and: