According to FDA's early alert (page current as of 23 September 2026), React Health stated that the internal flow transducer (IFT) cable connector in VOCSN ventilators with serial numbers 112786 and higher may partially or fully unseat, causing loss of communication with the IFT board. This can lead to reboots and recurring System Fault 10 alarms, and dropping the device can cause it. FDA said that a reboot interrupting therapy could delay therapy or cause a serious adverse event. The alert covers VOCSN+Pro, VC+Pro, VOCSN, VCSN+Pro, V+Pro, V+Pro Emergency, VC and V*Home, including Japan variants. FDA said no reboots during active ventilation had been reported, that it considers such a reboot during active therapy 'remote', and that as of 12 August 2026 React Health had reported no serious injuries or deaths.
React Health sent its customer letter on 17 September 2026. FDA's recall database lists the VOCSN+Pro (model PRT-00490-001) entry as a Class 1 recall, Z-0122-2027, posted 20 October 2026 and still open. It records 4,369 units in commerce, distributed across the US including Puerto Rico and outside the US to Japan and China; FDA lists the cause as under investigation by the firm. Recommended actions include using backup ventilation as needed, not stopping therapy unless an alternative is available and a clinician instructs it, keeping alternative ventilation equipment ready, continuously monitoring ventilator-dependent patients, with pulse oximetry and CO2 monitoring if available, and performing the checkout procedure after a drop or impact.[1,2,3]