According to New Zealand's Medsafe, the manufacturer identified the problem from customer complaints. During an active procedure for one patient, a user may manually select a different patient record from the 'Patient Cases' menu and then print images manually. The printout shows the correct image but may carry the selected patient's name, date of birth and medical record number instead of those of the patient in the procedure. Auto-print is not affected. FDA's recall entry (Z-0118-2027) says the firm began the recall on 23 September 2026 after a customer letter dated 9 September, and lists 58 systems distributed in the US and in Australia, Canada and New Zealand. FDA lists the root cause as software design. Users could keep operating the system before the correction but were told to confirm that manually printed images were correctly labelled before clinical use, filing or sharing. Medsafe lists the action as a software upgrade, starting 15 September 2026 in New Zealand.[1,2]