STERIS HexaVue Focus operating-room system can print another patient's identifiers on images

STERIS recalled HexaVue Focus Integration System software 2.0.18 (FDA Class 2). If a user selected a different patient record during a procedure and then printed manually, the printed images could carry that other patient's name, date of birth and medical record number.

Sources checked when written 28 November 2026

TierThe system stayed in use with a manual label check pending a software fix; wrong output does not change tier. How tiers are assigned.

What happened

According to New Zealand's Medsafe, the manufacturer identified the problem from customer complaints. During an active procedure for one patient, a user may manually select a different patient record from the 'Patient Cases' menu and then print images manually. The printout shows the correct image but may carry the selected patient's name, date of birth and medical record number instead of those of the patient in the procedure. Auto-print is not affected. FDA's recall entry (Z-0118-2027) says the firm began the recall on 23 September 2026 after a customer letter dated 9 September, and lists 58 systems distributed in the US and in Australia, Canada and New Zealand. FDA lists the root cause as software design. Users could keep operating the system before the correction but were told to confirm that manually printed images were correctly labelled before clinical use, filing or sharing. Medsafe lists the action as a software upgrade, starting 15 September 2026 in New Zealand.[1,2]

Documented harm

None documented.

What it teaches

  • Any output that carries patient identifiers should take them from the active case, not from whatever record the user is currently viewing.
  • Printed and exported clinical images can leave the system that created them, so identification errors on them can spread into other records.
  • When a fix is pending, an interim manual check of identifiers needs to be built into the workflow, not left to memory.

Sources

  1. Class 2 Device Recall: HexaVue Focus Integration System (Z-0118-2027). U.S. Food and Drug Administration, 7 October 2026. Primary Official report · link checked 2026-11-02
  2. Recall: HexaVue Focus Integration System (Medsafe ref 37551). Medsafe (New Zealand Medicines and Medical Devices Safety Authority), 15 September 2026. Primary Official report · link checked 2026-11-28

Study of 2017-2025 FDA data: neurosurgical software recalls, mostly Class I, with adverse-event reports continuing after recall

A peer-reviewed analysis in Neurosurgical Focus identified 20 software-associated neurological device recalls initiated in the US from 2017 to 2025, more than half of them FDA Class I. It counted 16,333 adverse-event notifications for these devices, more than 10,000 of them reported after formal recall notification for two devices.

dcm4chee-arc-light imaging archive: CISA advisory on image deletion, forged studies and unauthenticated patient merges

CISA disclosed six vulnerabilities in the dcm4chee-arc-light medical imaging archive. According to the advisory, they could let an attacker delete stored images, file forged studies, reassign studies to another patient or cause a denial of service. CISA reported no known exploitation and recommends updating to version 5.35.2 or later.

Luna G3 BPAP devices record the set pressure instead of the pressure delivered when Targeted VT is on; Class I recall

FDA classified a recall of the React Health (3B Medical) Luna G3 BPAP System, model LG3800-30VT, as Class 1. With Targeted VT enabled, the device records and transmits the configured inspiratory pressure instead of the pressure actually delivered. React Health says this could lead clinicians to adjust settings based on inaccurate data, which could delay therapy.

Tandem Mobi insulin pumps: false occlusion alarms suspend insulin delivery; Class 1 correction

Tandem Diabetes Care began a correction of Tandem Mobi insulin pumps in September 2026 because the pumps can raise false-positive occlusion alarms that suspend insulin delivery; FDA lists it as a Class 1 recall. A company spokesperson reported 230 adverse events requiring medical intervention or hospitalization as of 29 October 2026, and no deaths.

VOCSN ventilators: loose internal flow-transducer connector can cause reboots; FDA early alert and Class I recall

React Health (Ventec Life Systems) warned that the internal flow transducer cable connector in VOCSN ventilators can unseat and cause reboots and recurring System Fault 10 alarms, which could interrupt ventilation. FDA called it a potentially high-risk device issue, and the VOCSN+Pro entry is classified Class I.

Philips Lumify ultrasound demands online re-registration and blocks scanning in pre-hospital settings

FDA issued an Early Alert on 9 October 2026 and on 19 October 2026 said the issue had been classified as a Class I recall. Philips had reported that its Lumify ultrasound app may stop recognising a registered transducer and block scanning until the transducer is re-registered, which requires Wi-Fi or cellular access. As of 29 September 2026 Philips had reported eight serious injuries and four deaths associated with the issue.

PathConnectivity & data → Devices

All incidents · Devices

Information only, not advice. FailSystems is an aggregation and synthesis of published sources. It is not consulting, engineering, legal, regulatory or medical advice, and using it creates no professional relationship. Health systems are complex and no approach fits every organisation: anything you adopt is your own decision, at your own risk, and should be checked against the current official sources and by qualified people who know your setting. Full disclaimer.