Automated Impella Controllers recalled twice for purge cassette recognition failures; Abiomed reported 37 serious injuries and three deaths for the first issue

In 2026 Abiomed twice recalled its Automated Impella Controller, the console that runs Impella heart pumps, over failures to recognise the purge cassette. FDA classified both actions as Class I. FDA says Abiomed reported 37 serious injuries and three deaths linked to the first issue as of 17 August 2026.

Sources checked when written 28 November 2026

TierSources describe failures in individual controllers, managed by reinserting the purge disc or switching to a backup controller; no facility-level drop in care mode is documented. How tiers are assigned.

What happened

According to FDA, a letter from Abiomed dated 12 August 2026 told customers that the purge flag component in the controller's purge pressure sensor assembly could fail, causing purge cassette recognition problems. These showed up as "Purge Disc Disconnected" alarms at setup and "Purge Disc not Detected" alarms during support. FDA says Abiomed recommended removing all Automated Impella Controllers in phases, with hospitals using their current inventory while waiting for service. If an alarm did not clear after the purge disc was reinserted, users were told to switch to a backup controller. Abiomed plans to replace the purge flag and retainer components. FDA classified this as a Class I recall on 24 September 2026. A Johnson & Johnson spokesperson told TCTMD that the issue had been identified in 0.2% of cases. Abiomed sent a second removal letter on 18 September 2026 after a review of servicing records found units that needed hardware, firmware or software updates. FDA's recall record lists four problems. An RFID firmware issue may stop the controller recognising the purge cassette during setup or a cassette change, which can delay a procedure or therapy. The rotary push button may stick, which can lead to loss of control of the console. The controller may become unstable, which can cause start-up failure and an inability to drive the pump. A calculator software defect can produce negative central venous pressure values in pulmonary artery pressure calculations. The record lists 428 units distributed worldwide, with the cause still under investigation by the firm. FDA's notice says the updates were available "but have not been implemented consistently across all potentially impacted units". FDA identified this action as Class I on 9 November 2026. Separately, MedTech Dive reported a Class I correction in April 2026 for four hardware issues that could lead to a failure to boot the controller or a sudden interruption in haemodynamic support.[1,2,3,4,5]

Documented harm

According to FDA, as of 17 August 2026 Abiomed had reported 37 serious injuries and three deaths associated with the purge flag issue. For the September 2026 recall, FDA says Abiomed had reported no serious injuries or deaths as of 8 September 2026.

What it teaches

  • Life-sustaining consoles whose recognition of consumables can fail need an immediately available backup unit and a practised changeover.
  • Fleet owners should check that issued field updates have actually been applied to every unit, because updates that exist but are unevenly installed leave latent hazards in service.
  • Retrospective review of servicing records can surface latent hardware, firmware and software faults that individual service events do not reveal.

Sources

  1. Heart Pump Controller Recall: Abiomed Removes Automated Impella Controllers. U.S. Food and Drug Administration, 24 September 2026. Primary Official report · link checked 2026-11-17
  2. Heart Pump Controller Recall: Abiomed Removes Automated Impella Controllers (AIC). U.S. Food and Drug Administration, 9 November 2026. Primary Official report · link checked 2026-11-17
  3. Class 1 Device Recall Automated Impella Controller (Z-0208-2027). U.S. Food and Drug Administration, Medical Device Recalls database, 28 October 2026. Primary Dataset · link checked 2026-11-17
  4. Abiomed Identifies Purge-Control Issue With Impella: FDA. TCTMD, 27 August 2026. Supporting Journalism · link checked 2026-11-28
  5. J&J recalls Impella controllers over hardware issues. MedTech Dive, 12 June 2026. Secondary Journalism · link checked 2026-11-28

Study of 2017-2025 FDA data: neurosurgical software recalls, mostly Class I, with adverse-event reports continuing after recall

A peer-reviewed analysis in Neurosurgical Focus identified 20 software-associated neurological device recalls initiated in the US from 2017 to 2025, more than half of them FDA Class I. It counted 16,333 adverse-event notifications for these devices, more than 10,000 of them reported after formal recall notification for two devices.

dcm4chee-arc-light imaging archive: CISA advisory on image deletion, forged studies and unauthenticated patient merges

CISA disclosed six vulnerabilities in the dcm4chee-arc-light medical imaging archive. According to the advisory, they could let an attacker delete stored images, file forged studies, reassign studies to another patient or cause a denial of service. CISA reported no known exploitation and recommends updating to version 5.35.2 or later.

Luna G3 BPAP devices record the set pressure instead of the pressure delivered when Targeted VT is on; Class I recall

FDA classified a recall of the React Health (3B Medical) Luna G3 BPAP System, model LG3800-30VT, as Class 1. With Targeted VT enabled, the device records and transmits the configured inspiratory pressure instead of the pressure actually delivered. React Health says this could lead clinicians to adjust settings based on inaccurate data, which could delay therapy.

Tandem Mobi insulin pumps: false occlusion alarms suspend insulin delivery; Class 1 correction

Tandem Diabetes Care began a correction of Tandem Mobi insulin pumps in September 2026 because the pumps can raise false-positive occlusion alarms that suspend insulin delivery; FDA lists it as a Class 1 recall. A company spokesperson reported 230 adverse events requiring medical intervention or hospitalization as of 29 October 2026, and no deaths.

VOCSN ventilators: loose internal flow-transducer connector can cause reboots; FDA early alert and Class I recall

React Health (Ventec Life Systems) warned that the internal flow transducer cable connector in VOCSN ventilators can unseat and cause reboots and recurring System Fault 10 alarms, which could interrupt ventilation. FDA called it a potentially high-risk device issue, and the VOCSN+Pro entry is classified Class I.

All incidents · Devices

Information only, not advice. FailSystems is an aggregation and synthesis of published sources. It is not consulting, engineering, legal, regulatory or medical advice, and using it creates no professional relationship. Health systems are complex and no approach fits every organisation: anything you adopt is your own decision, at your own risk, and should be checked against the current official sources and by qualified people who know your setting. Full disclaimer.