On 22 July 2025 FDA issued an early alert about Baxter's Novum IQ large volume pump (LVP). According to the FDA notice, Baxter reported that the pump can underinfuse when it moves from one flow rate to a rate more than double the first, for example during a rate change or bolus, and that in the worst case no delivery may occur. Baxter also reported more customer complaints of over- and underinfusion, possibly caused by tubing misloaded into the pump channel. Baxter's 14 July letter told customers to change the administration set, or switch to another pump with a new set, at the first safe opportunity. Where that delay was unacceptable, staff were to move the set about 0.5 inches downstream before starting a bolus or a rate change above 100%. FDA classified this correction as a Class I recall on 22 December 2025.
A second correction, in a Baxter letter dated 4 August 2025, covered both the Novum IQ LVP and the Novum IQ syringe pump. According to FDA's notice, software anomalies can leave a blank Run screen: therapy continues, but it cannot be adjusted without interrupting it. On syringe pumps, a false motor movement system error (codes 20602 or 20603) can interrupt therapy after certain dose or rate changes below the minimum recommended flow rate on syringes of 10 mL or larger. Baxter advised customers to consider an alternate pump for high-risk medications or critically ill patients, if one is available and appropriate. In its Form 10-Q for the quarter ended 30 September 2026, Baxter states three things: FDA classified its 2025 Novum LVP corrections as Class I recalls; in July 2025 it elected to temporarily stop distributing and installing the Novum LVP in the US and Canada, except in cases of medical necessity; and some further corrections may require regulatory clearance or approval and are in the early stages of verification testing. Investors filed putative securities class actions in the Northern District of Illinois in October and December 2025. According to the 10-Q, these allege false and misleading statements and a failure to disclose material facts relating to the Novum LVP. A Baxter spokesperson told Drug Delivery Business that the company does not comment on pending litigation and intends to vigorously defend its position.[1,2,3,4,5]