FDA's notice lists an Early Alert on 15 July 2025 for certain BD Alaris pump infusion sets used with the Alaris Pump Module model 8100 and classification as a Class I recall on 29 September 2025; Medical Daily gives 17 July 2025 for the Early Alert and 29 September 2026 for the classification. In a September 2025 update published by FDA, BD added 15 more sets to the recall, said it had received no complaints associated with the issue, and noted a higher risk to vulnerable patients such as neonates and the critically ill.
On 19 August 2026 BD and CareFusion 303 sent customers an urgent medical device correction. FDA's recall record Z-3080-2026 gives the reason as 'discrepancies in specific attributes in the performance data for infusion sets, which may lead to inappropriate pump performance', a cause of device design, and 94,230,757 units in commerce; Medical Daily reported that FDA's database lists 33,162,680 affected units, and the sources do not reconcile the two figures. In FDA's 5 October 2026 update, BD stated that some sets with restrictive components and designs 'perform worse than previously disclosed', most commonly at programmed rates under 1 mL/h and bolus volumes of 5 mL or less, and that previously claimed performance was overestimated. FDA relays BD's advice to discard listed discontinued sets, prioritise alternative sets in critical-care, neonatal, infant and paediatric areas, and use a syringe pump where set performance is not clinically appropriate. Medical Daily reported that BD had not publicly explained how the discrepancy was found beyond continued testing.[1,2,3,4,5,6]