Incident

LeadCare blood lead analyzers return falsely low results; malfunction concealed

Magellan's LeadCare blood lead analyzers gave falsely low results on venous samples. LeadCare II alone accounted for more than half of US blood lead tests in 2013-2017. The company reported the LeadCare Ultra malfunction to FDA nearly 21 months after discovering it.

Fact-checked 11 November 2026

TierThe automation stayed up and kept producing output that was wrong or went unchecked; there was no outage to fall back from. How tiers are assigned.

What happened

Magellan Diagnostics first learned during the FDA clearance process in June 2013 that LeadCare Ultra could give inaccurate lead results. It released the device in late 2013 without informing customers or FDA. It did not report the malfunction to FDA until April 2015: nearly 21 months after discovering it, and almost 8 months after customers found it on their own in August 2014. It did not notify FDA about the LeadCare II malfunction until November 2016. It admitted changing the devices' user instructions without prior FDA notice or approval. According to DOJ, the malfunction produced inaccurately low lead results for tens of thousands of children and other patients. In June 2024 the company pleaded guilty to two counts of introducing a misbranded medical device into interstate commerce under the FD&C Act; the plea agreement provided for a $28.1 million fine. At sentencing in October 2024 it was ordered to pay a $21.8 million fine, $10.9 million in forfeiture and a minimum of $9.3 million to compensate patient victims.[1,2]

Documented harm

Tens of thousands of children and other patients received inaccurately low lead results (DOJ); individual clinical outcomes not quantified.

What it teaches

  • A lab analyzer's wrong number is indistinguishable from a right one without an independent method comparison.
  • Manufacturer reporting is a control that can fail through concealment; your own proficiency testing and discrepancy tracking is the backstop.
  • Point-of-care tests that dominate a national testing market concentrate error at population scale.

Sources

  1. Magellan Diagnostics Pleads Guilty to Criminal FDCA Charges. U.S. Department of Justice (District of Massachusetts), reposted by FDA OCI, 27 June 2024. Primary Official report · link checked 2026-11-02
  2. Magellan Diagnostics Sentenced for Concealing Malfunction in Lead Testing Devices. U.S. Department of Justice (District of Massachusetts), reposted by FDA OCI, 10 October 2024. Primary Official report · link checked 2026-11-28

Study of 2017-2025 FDA data: neurosurgical software recalls, mostly Class I, with adverse-event reports continuing after recall

A peer-reviewed analysis in Neurosurgical Focus identified 20 software-associated neurological device recalls initiated in the US from 2017 to 2025, more than half of them FDA Class I. It counted 16,333 adverse-event notifications for these devices, more than 10,000 of them reported after formal recall notification for two devices.

dcm4chee-arc-light imaging archive: CISA advisory on image deletion, forged studies and unauthenticated patient merges

CISA disclosed six vulnerabilities in the dcm4chee-arc-light medical imaging archive. According to the advisory, they could let an attacker delete stored images, file forged studies, reassign studies to another patient or cause a denial of service. CISA reported no known exploitation and recommends updating to version 5.35.2 or later.

Luna G3 BPAP devices record the set pressure instead of the pressure delivered when Targeted VT is on; Class I recall

FDA classified a recall of the React Health (3B Medical) Luna G3 BPAP System, model LG3800-30VT, as Class 1. With Targeted VT enabled, the device records and transmits the configured inspiratory pressure instead of the pressure actually delivered. React Health says this could lead clinicians to adjust settings based on inaccurate data, which could delay therapy.

Tandem Mobi insulin pumps: false occlusion alarms suspend insulin delivery; Class 1 correction

Tandem Diabetes Care began a correction of Tandem Mobi insulin pumps in September 2026 because the pumps can raise false-positive occlusion alarms that suspend insulin delivery; FDA lists it as a Class 1 recall. A company spokesperson reported 230 adverse events requiring medical intervention or hospitalization as of 29 October 2026, and no deaths.

VOCSN ventilators: loose internal flow-transducer connector can cause reboots; FDA early alert and Class I recall

React Health (Ventec Life Systems) warned that the internal flow transducer cable connector in VOCSN ventilators can unseat and cause reboots and recurring System Fault 10 alarms, which could interrupt ventilation. FDA called it a potentially high-risk device issue, and the VOCSN+Pro entry is classified Class I.

All incidents · Devices

Information only, not advice. FailSystems is an aggregation and synthesis of published sources. It is not consulting, engineering, legal, regulatory or medical advice, and using it creates no professional relationship. Health systems are complex and no approach fits every organisation: anything you adopt is your own decision, at your own risk, and should be checked against the current official sources and by qualified people who know your setting. Full disclaimer.