StatStrip Gen 2.0 hospital glucose meters can file a result under the previously scanned patient; Class I recall

Nova Biomedical's StatStrip Gen 2.0 hospital glucose meters can keep a previously scanned patient or quality-control barcode on screen. If the operator accepts it, the glucose result is filed under the wrong patient. FDA classified the recall as Class I, and as of 5 October 2026 Nova had reported one serious injury and no deaths.

Sources checked when written 4 November 2026

TierThe meters stayed in use with added manual identity checks, or were swapped for alternative meters; wrong output does not change tier, and no loss of automation is documented. How tiers are assigned.

What happened

FDA's notice, issued on 20 October 2026 and updated on 30 October to clarify recommended actions, says earlier information can stay on the meter screen when patient ID or quality-control barcodes are scanned. If the operator accepts it, the glucose result is assigned to the wrong patient's record or QC ID. Nova says analytical performance is not affected and that the issue affects all firmware versions of the Gen 2.0 model (models 63685 and 66739). Nova's announcement says earlier generations (1.0, 1.75, 1.86 and 1.87) are not affected. According to FDA, clinicians may act on another patient's result, which can lead to delayed treatment, unnecessary treatment including inappropriate insulin, or no treatment for the intended patient. FDA says critically ill patients and infants are at greater risk. Nova's announcement, posted by FDA on 6 October 2026, says the correction began on 24 July 2026. It says that on 17 July 2026 Nova received one U.S. report of inappropriate insulin administration based on another patient's glucose result, and it also states that no serious adverse events related to the issue have been reported. FDA's notice says that as of 5 October Nova had reported one serious injury and no deaths. FDA dates Nova's first customer letter 31 July, with an update on 5 October. Nova has put new Gen 2.0 shipments on hold and plans a phased software update starting in November 2026. Until then, FDA's notice tells users to enable the 'Validate Patient ID's' setting in NovaNet where ADT data is used, and to set operator logoff to 'Test' mode. Staff should check the patient ID at scanning and again on the result screen, and flag or remove any misassigned result and tell the administrator. Facilities with an alternative meter should stop using the affected meters until they are updated. Clinical Lab Products reported the same defect, injury count and interim measures.[1,2,3]

Documented harm

FDA's notice says that as of 5 October 2026 Nova had reported one serious injury and no deaths. Nova's announcement describes one U.S. report of inappropriate insulin administration based on another patient's result, while also stating that no serious adverse events related to the issue had been reported.

What it teaches

  • Barcode patient identification only prevents wrong-patient errors if the device clears the previous identity before it accepts a new scan.
  • Checking device-captured IDs against admission (ADT) data can catch mismatches that a busy operator accepts.
  • A defect that leaves measurement accuracy intact can still cause treatment errors when the result is attached to the wrong person.

Sources

  1. Hospital Glucose Meter Correction: Nova Biomedical Updates Use Instructions for StatStrip Glucose Hospital Meter Systems. U.S. Food and Drug Administration, 20 October 2026. Primary Regulation · link checked 2026-11-04
  2. Nova Biomedical Announces Global Medical Device Correction for StatStrip Glucose Hospital Meter. Nova Biomedical (company announcement posted by FDA), 6 October 2026. Supporting Official report · link checked 2026-11-02
  3. Nova Biomedical Issues Class I Recall for StatStrip Glucose Meters. Clinical Lab Products, 2 November 2026. Supporting Journalism · link checked 2026-11-04

Study of 2017-2025 FDA data: neurosurgical software recalls, mostly Class I, with adverse-event reports continuing after recall

A peer-reviewed analysis in Neurosurgical Focus identified 20 software-associated neurological device recalls initiated in the US from 2017 to 2025, more than half of them FDA Class I. It counted 16,333 adverse-event notifications for these devices, more than 10,000 of them reported after formal recall notification for two devices.

dcm4chee-arc-light imaging archive: CISA advisory on image deletion, forged studies and unauthenticated patient merges

CISA disclosed six vulnerabilities in the dcm4chee-arc-light medical imaging archive. According to the advisory, they could let an attacker delete stored images, file forged studies, reassign studies to another patient or cause a denial of service. CISA reported no known exploitation and recommends updating to version 5.35.2 or later.

Luna G3 BPAP devices record the set pressure instead of the pressure delivered when Targeted VT is on; Class I recall

FDA classified a recall of the React Health (3B Medical) Luna G3 BPAP System, model LG3800-30VT, as Class 1. With Targeted VT enabled, the device records and transmits the configured inspiratory pressure instead of the pressure actually delivered. React Health says this could lead clinicians to adjust settings based on inaccurate data, which could delay therapy.

Tandem Mobi insulin pumps: false occlusion alarms suspend insulin delivery; Class 1 correction

Tandem Diabetes Care began a correction of Tandem Mobi insulin pumps in September 2026 because the pumps can raise false-positive occlusion alarms that suspend insulin delivery; FDA lists it as a Class 1 recall. A company spokesperson reported 230 adverse events requiring medical intervention or hospitalization as of 29 October 2026, and no deaths.

VOCSN ventilators: loose internal flow-transducer connector can cause reboots; FDA early alert and Class I recall

React Health (Ventec Life Systems) warned that the internal flow transducer cable connector in VOCSN ventilators can unseat and cause reboots and recurring System Fault 10 alarms, which could interrupt ventilation. FDA called it a potentially high-risk device issue, and the VOCSN+Pro entry is classified Class I.

All incidents · Devices

Information only, not advice. FailSystems is an aggregation and synthesis of published sources. It is not consulting, engineering, legal, regulatory or medical advice, and using it creates no professional relationship. Health systems are complex and no approach fits every organisation: anything you adopt is your own decision, at your own risk, and should be checked against the current official sources and by qualified people who know your setting. Full disclaimer.