Philips Respironics ventilator, BiPAP and CPAP recall over degrading sound-abatement foam

Philips recalled about 15 million breathing devices because PE-PUR foam could degrade into particles and chemicals the patient could inhale; remediation ran years and ended in a consent decree.

Sources checked when written 9 November 2026

TierNo loss of automation is documented. How tiers are assigned.

What happened

On 30 June 2021 Philips Respironics recalled certain ventilators, BiPAP and CPAP machines (Class I) because the polyester-based polyurethane foam used to reduce noise could break down, releasing particles or invisible chemicals into the air path. FDA issued a section 518(a) order in March 2022 over the adequacy of recall notifications. By 31 January 2024 FDA had received more than 116,000 MDRs, including 561 death reports, associated with foam breakdown; FDA notes MDRs cannot by themselves establish cause. On 9 April 2024 a federal court entered a consent decree restricting new device production at three Philips facilities until compliance, requiring a recall remediation plan and independent review of testing of the replacement silicone foam. Separately, in July 2024 a Class I software correction for Trilogy Evo ventilators addressed issues including an inaccurate display of delivered oxygen.[1,2,3,4]

Documented harm

561 death reports and over 116,000 MDRs associated with foam breakdown as of Jan 2024; causation not established by MDRs alone.

What it teaches

  • A hazard can sit inside a working device for years with no alarm or visible symptom.
  • Recall execution (replacement supply, notification, patient tracing) is a failure point of its own.
  • Home-use life-support devices need a registry the hospital or DME supplier can query when a recall lands.

Sources

  1. FDA Activities Related to Recalled Philips Ventilators, BiPAP Machines, and CPAP Machines. U.S. Food and Drug Administration, November 2024. Primary Official report · link checked 2026-11-02
  2. Problems Reported with Recalled Philips Ventilators, BiPAP Machines, and CPAP Machines. U.S. Food and Drug Administration, 31 January 2024. Primary Dataset · link checked 2026-11-02
  3. Ventilator Software Correction: Philips Respironics Issues Mandatory Software Correction and Updates Use Instructions for Trilogy Evo, EV300, EvoO2, and Evo Universal. U.S. Food and Drug Administration, 16 July 2024. Primary Official report · link checked 2026-11-02

Study of 2017-2025 FDA data: neurosurgical software recalls, mostly Class I, with adverse-event reports continuing after recall

A peer-reviewed analysis in Neurosurgical Focus identified 20 software-associated neurological device recalls initiated in the US from 2017 to 2025, more than half of them FDA Class I. It counted 16,333 adverse-event notifications for these devices, more than 10,000 of them reported after formal recall notification for two devices.

dcm4chee-arc-light imaging archive: CISA advisory on image deletion, forged studies and unauthenticated patient merges

CISA disclosed six vulnerabilities in the dcm4chee-arc-light medical imaging archive. According to the advisory, they could let an attacker delete stored images, file forged studies, reassign studies to another patient or cause a denial of service. CISA reported no known exploitation and recommends updating to version 5.35.2 or later.

Luna G3 BPAP devices record the set pressure instead of the pressure delivered when Targeted VT is on; Class I recall

FDA classified a recall of the React Health (3B Medical) Luna G3 BPAP System, model LG3800-30VT, as Class 1. With Targeted VT enabled, the device records and transmits the configured inspiratory pressure instead of the pressure actually delivered. React Health says this could lead clinicians to adjust settings based on inaccurate data, which could delay therapy.

Tandem Mobi insulin pumps: false occlusion alarms suspend insulin delivery; Class 1 correction

Tandem Diabetes Care began a correction of Tandem Mobi insulin pumps in September 2026 because the pumps can raise false-positive occlusion alarms that suspend insulin delivery; FDA lists it as a Class 1 recall. A company spokesperson reported 230 adverse events requiring medical intervention or hospitalization as of 29 October 2026, and no deaths.

VOCSN ventilators: loose internal flow-transducer connector can cause reboots; FDA early alert and Class I recall

React Health (Ventec Life Systems) warned that the internal flow transducer cable connector in VOCSN ventilators can unseat and cause reboots and recurring System Fault 10 alarms, which could interrupt ventilation. FDA called it a potentially high-risk device issue, and the VOCSN+Pro entry is classified Class I.

All incidents · Devices

Information only, not advice. FailSystems is an aggregation and synthesis of published sources. It is not consulting, engineering, legal, regulatory or medical advice, and using it creates no professional relationship. Health systems are complex and no approach fits every organisation: anything you adopt is your own decision, at your own risk, and should be checked against the current official sources and by qualified people who know your setting. Full disclaimer.