Pulse oximeters overestimate oxygen saturation in patients with darker skin

Paired SpO2/SaO2 data showed occult hypoxemia missed by pulse oximetry about three times as often in Black as in White patients, delaying treatment decisions.

Fact-checked 25 November 2026

What happened

Sjoding and colleagues compared pulse oximetry with arterial blood gas measurements. Occult hypoxemia (SaO2 below 88% while SpO2 read 92-96%) occurred in 11.7% of measurements in Black patients vs 3.6% in White patients at Michigan, and 17.0% vs 6.2% in a multicenter cohort. A 2022 Johns Hopkins COVID-19 study found SpO2 overestimated SaO2 by 1.1-1.7 points in Asian, Black and Hispanic patients relative to White patients, and Black patients had a 29% lower hazard of being recognized as eligible for oxygen-threshold therapies. FDA issued a safety communication in February 2021, held advisory panels in November 2022 and February 2024, and issued draft guidance on 7 January 2025 revising performance testing across skin pigmentation. As of this review FDA's product page still lists the January 2025 guidance as draft.[1,2,3,4]

Documented harm

Documented delayed or missed recognition of treatment eligibility (Fawzy 2022); individual patient outcomes not attributed.

What it teaches

  • A device that keeps reporting a plausible wrong number triggers no downtime procedure at all.
  • Validation populations define who the device is accurate for; ask for the data by skin tone.
  • Thresholds built on a biased sensor (triage, therapy eligibility, early-warning scores) inherit the bias.

Sources

  1. Racial Bias in Pulse Oximetry Measurement. New England Journal of Medicine (Sjoding MW, Dickson RP, Iwashyna TJ, Gay SE, Valley TS), 17 December 2020. Primary Peer-reviewed · link checked 2026-11-02
  2. Racial and Ethnic Discrepancy in Pulse Oximetry and Delayed Identification of Treatment Eligibility Among Patients With COVID-19. JAMA Internal Medicine (Fawzy A, et al.), 1 July 2022. Primary Peer-reviewed · link checked 2026-11-02
  3. Pulse Oximeters (FDA actions on accuracy and skin pigmentation). U.S. Food and Drug Administration, 6 January 2025. Primary Guidance · link checked 2026-11-02
  4. Pulse Oximeters for Medical Purposes - Non-Clinical and Clinical Performance Testing, Labeling, and Premarket Submission Recommendations (Draft Guidance). U.S. Food and Drug Administration, 7 January 2025. Primary Guidance · link checked 2026-11-02

Study of 2017-2025 FDA data: neurosurgical software recalls, mostly Class I, with adverse-event reports continuing after recall

A peer-reviewed analysis in Neurosurgical Focus identified 20 software-associated neurological device recalls initiated in the US from 2017 to 2025, more than half of them FDA Class I. It counted 16,333 adverse-event notifications for these devices, more than 10,000 of them reported after formal recall notification for two devices.

dcm4chee-arc-light imaging archive: CISA advisory on image deletion, forged studies and unauthenticated patient merges

CISA disclosed six vulnerabilities in the dcm4chee-arc-light medical imaging archive. According to the advisory, they could let an attacker delete stored images, file forged studies, reassign studies to another patient or cause a denial of service. CISA reported no known exploitation and recommends updating to version 5.35.2 or later.

Luna G3 BPAP devices record the set pressure instead of the pressure delivered when Targeted VT is on; Class I recall

FDA classified a recall of the React Health (3B Medical) Luna G3 BPAP System, model LG3800-30VT, as Class 1. With Targeted VT enabled, the device records and transmits the configured inspiratory pressure instead of the pressure actually delivered. React Health says this could lead clinicians to adjust settings based on inaccurate data, which could delay therapy.

Tandem Mobi insulin pumps: false occlusion alarms suspend insulin delivery; Class 1 correction

Tandem Diabetes Care began a correction of Tandem Mobi insulin pumps in September 2026 because the pumps can raise false-positive occlusion alarms that suspend insulin delivery; FDA lists it as a Class 1 recall. A company spokesperson reported 230 adverse events requiring medical intervention or hospitalization as of 29 October 2026, and no deaths.

VOCSN ventilators: loose internal flow-transducer connector can cause reboots; FDA early alert and Class I recall

React Health (Ventec Life Systems) warned that the internal flow transducer cable connector in VOCSN ventilators can unseat and cause reboots and recurring System Fault 10 alarms, which could interrupt ventilation. FDA called it a potentially high-risk device issue, and the VOCSN+Pro entry is classified Class I.

All incidents · Devices · Models & agents

Information only, not advice. FailSystems is an aggregation and synthesis of published sources. It is not consulting, engineering, legal, regulatory or medical advice, and using it creates no professional relationship. Health systems are complex and no approach fits every organisation: anything you adopt is your own decision, at your own risk, and should be checked against the current official sources and by qualified people who know your setting. Full disclaimer.