Incident

Spacelabs Xhibit Central Station: DisplayPort splitter can cause loss of alarm audio and video; Class 2 correction

On 4 June 2026 Spacelabs Healthcare told customers to stop using a DisplayPort splitter (PN 010-1975-00) with its Xhibit Central Station patient-monitoring system. According to the company's field safety notice, the splitter can cause degraded or complete loss of alarm audio and video at the central station or extended displays. FDA lists the action as a Class 2 recall (Z-2916-2026).

Sources checked when written 28 November 2026

TierThe sources describe a potential loss of central-station alarm audio and video and document no event in which care dropped to a lower tier. How tiers are assigned.

What happened

The Xhibit Central Station (model 96102) gives clinicians central monitoring of data from networked Spacelabs patient monitors and telemetry transmitters. It also generates visible and audible alarms. A four-way DisplayPort splitter copies the station's display and audio to as many as three remote displays, for example in corridors, care alcoves or break rooms. FDA's recall entry describes the splitter as a third-party external accessory. It gives the manufacturer's reason for recall as an "off-the-shelf display port component that has audio/video failures" and lists the FDA-determined cause as "Unknown/Undetermined by firm". According to Spacelabs' field safety notice, the problem was found through a review of product complaints about lost video and audio. The notice says splitters "may fail to perform as intended and result in degraded or complete loss of alarm audio and video" and that unqualified splitters may pose the same risks. It states that clinical staff may then be unable to detect an alarm condition and may fail to respond to a medical condition that requires intervention, and that serious injury or death is foreseeable. On 4 June 2026 Spacelabs emailed customers an urgent medical device correction letter. The letter told them to stop using splitter 010-1975-00 or any unapproved splitter, notify staff, acknowledge the notice, dispose of the affected components, and install a newly qualified replacement (model 010-1977-00, kit 020-0235-00). According to the French-language notice published by ANSM, customers who bought the affected splitter from Spacelabs receive the replacement free of charge. The English notice published by the Dutch inspectorate IGJ sets 30 August 2026 as the completion date. FDA lists 6,074 units in commerce, US nationwide distribution and several other countries. It posted the recall on 7 August 2026, and the recall remained open.[1,2,3]

Documented harm

None documented. The notice published by ANSM states that no injury or death had been reported to date.

What it teaches

  • Commercial accessories that carry alarm audio or video to remote displays belong to the alarm system and need the same qualification and change control as the monitor itself.
  • Remote displays that copy central-station alarms can fail without anyone noticing, so periodic checks that alarms are audible and visible at each display location can find such failures before an alarm is missed.
  • Relying on one path to deliver alarm audio leaves no backup when a component in that path fails; a separate notification channel gives clinicians a second way to learn of an alarm.

Sources

  1. Class 2 Device Recall Xhibit (Z-2916-2026). U.S. Food and Drug Administration, Medical Device Recalls database, 7 August 2026. Primary Official report · link checked 2026-11-10
  2. Spacelabs Healthcare Ltd-3010157426-05042026-003-C-Xhibit Central Station (Urgent Field Safety Notice: Audio/Video Loss With Splitter use). Inspectie Gezondheidszorg en Jeugd (IGJ), publishing Spacelabs Healthcare's field safety notice, 25 June 2026. Primary Official report · link checked 2026-11-28
  3. Correction urgente d'un dispositif médical: Station centrale Xhibit utilisée avec un répartiteur DisplayPort – Perte de signaux audio et/ou vidéo. ANSM (Agence nationale de sécurité du médicament et des produits de santé), publishing Spacelabs Healthcare's notice, 10 June 2026. Primary Official report · link checked 2026-11-28

Study of 2017-2025 FDA data: neurosurgical software recalls, mostly Class I, with adverse-event reports continuing after recall

A peer-reviewed analysis in Neurosurgical Focus identified 20 software-associated neurological device recalls initiated in the US from 2017 to 2025, more than half of them FDA Class I. It counted 16,333 adverse-event notifications for these devices, more than 10,000 of them reported after formal recall notification for two devices.

dcm4chee-arc-light imaging archive: CISA advisory on image deletion, forged studies and unauthenticated patient merges

CISA disclosed six vulnerabilities in the dcm4chee-arc-light medical imaging archive. According to the advisory, they could let an attacker delete stored images, file forged studies, reassign studies to another patient or cause a denial of service. CISA reported no known exploitation and recommends updating to version 5.35.2 or later.

Luna G3 BPAP devices record the set pressure instead of the pressure delivered when Targeted VT is on; Class I recall

FDA classified a recall of the React Health (3B Medical) Luna G3 BPAP System, model LG3800-30VT, as Class 1. With Targeted VT enabled, the device records and transmits the configured inspiratory pressure instead of the pressure actually delivered. React Health says this could lead clinicians to adjust settings based on inaccurate data, which could delay therapy.

Tandem Mobi insulin pumps: false occlusion alarms suspend insulin delivery; Class 1 correction

Tandem Diabetes Care began a correction of Tandem Mobi insulin pumps in September 2026 because the pumps can raise false-positive occlusion alarms that suspend insulin delivery; FDA lists it as a Class 1 recall. A company spokesperson reported 230 adverse events requiring medical intervention or hospitalization as of 29 October 2026, and no deaths.

VOCSN ventilators: loose internal flow-transducer connector can cause reboots; FDA early alert and Class I recall

React Health (Ventec Life Systems) warned that the internal flow transducer cable connector in VOCSN ventilators can unseat and cause reboots and recurring System Fault 10 alarms, which could interrupt ventilation. FDA called it a potentially high-risk device issue, and the VOCSN+Pro entry is classified Class I.

All incidents · Devices

Information only, not advice. FailSystems is an aggregation and synthesis of published sources. It is not consulting, engineering, legal, regulatory or medical advice, and using it creates no professional relationship. Health systems are complex and no approach fits every organisation: anything you adopt is your own decision, at your own risk, and should be checked against the current official sources and by qualified people who know your setting. Full disclaimer.